PURELAB Ultra Mk 2
Operator Manual
PURELAB Ultra Mk2 Version 3 11/08
Page 15
5.3 Connecting
the
PURELAB Ultra
Once the
PURELAB Ultra
unit has been positioned either on a wall or
on a bench, it should be connected as follows:
•
Mains water inlet tube
•
Drain tube - to drain
•
Flush tube (Bioscience and Genetic) - to drain
Step 1 - Remove Transit Plugs
1. PUSH in collet on connector.
2. PULL out transit plug.
3. CUT a clean square end on an 8 mm OD semi rigid
drain tube.
4. PUSH tube into connector.
Step 2 - Connect Water Inlet
1. Connect 8 mm tube from the
PURELAB Ultra
inlet to
the inlet strainer.
2. Connect the inlet strainer to the pre-purified water
supply.
Step 3 - Connect Drains
1. Connect 8 mm tube from the drain and UF drain (if
fitted) and direct to a suitable drainage point.
CAUTION!
The drain and flush lines should allow a
gravity fall to drain with no restrictions.
UF Drain
Drain
Inlet
Inlet and Drain Connection
Strainer
Inlet
Collar
Collar
Direction of
water flow
Mesh filter
Feed Water Inlet
Connector
Collet
Blanking
transit plug
Pull plug
Depress
collet
Clean square
cut end
Tube 8mm OD
Simply push in tube to attach
Tube in secured position
Fitting tubes