6.2 Manual
Dispense
Step 1 - Product Quality
1. ENSURE
PURELAB Ultra
is in recirculation mode and
the water quality is satisfactory.
Step 2 - Dispense
1. PRESS Dispense Controller once.
2. ROTATE Dispense Controller clockwise.
If the Dispense Controller is not rotated within 15
seconds the unit will revert to recirculation mode.
3. ADJUST the flow from the unit by rotating the Dispense
Controller as necessary.
Note: To prevent accidentally leaving the unit on
continuous droplet dispense the
PURELAB
Ultra
will only dispense for a maximum of 4
minutes from the last turn of the Dispense
Controller. For large volumes it is
recommended that the Volumetric Dispense
mode be used. See Section 6.3 - Volumetric
Dispense.
4. PRESS Dispense Controller once to stop dispensing
and return to recirculation mode.
6.3 Profile
Dispense
The PURELAB Ultra includes the ability to record and replay the
sequence of events which occur during manual dispense. This feature
allows you to dispense non-standard volumes in a precise and
repeatable manner.
6.4 Dispense
Profiling
Step 1 - Record Dispense Profile
1. ENSURE unit is in recirculation mode and displaying
adequate water quality.
2. PRESS dispense controller once. The screen now
displays three M+ icons above the soft buttons.
3. PRESS button into which you would like to record the
profile.
4. ROTATE Dispense controller and dispense. During this
period it is possible to vary the flow from zero through to
maximum or as required.
5. PRESS dispense controller to stop the dispense and
save the profile.
Manual Dispense Screen