8. SANITIZATION
PROCEDURES
The unit is sanitized to destroy the bacteria within the pipework, and
the filters of the unit. Please read this entire section to become familiar
with the procedure before you start. Sanitization may be required in
the following circumstances:
•
Once a month to maintain low bacterial counts
•
If the unit has not been used for a prolonged period of time
•
If the product water is used for particularly stringent
applications
Step 1 - Start Sanitization Cycle
1. ENSURE unit is ready to use.
2. SWITCH OFF electrical supply.
3. REMOVE POU filter, if fitted.
4. OPEN the door.
Step 2 - Insert Tablet into One Sanitization By-pass Block Only
1. REMOVE one of the two by-pass blocks from the inside
of the door.
2. INVERT the by-pass block.
3. UNSCREW cap on sanitization by-pass block.
4. INSERT CT1 cleaning tablet.
5. REFIT cap on by-pass block, hand tight.
Step 3 - Remove Both Purification Packs
1. REMOVE Polishing Purification Pack. (See section 7.1).
2. REMOVE Primary Purification Pack. (See section 7.1).
3. PLACE Purification Packs safely to one side.
Screens
By-pass blocks
Primary
Polishing
Purification
Purification
Packs
Packs
Location of By-pass Blocks inside Door
Cap
CT1 cleaning
tablet
Insert CT1 Cleaning Tablet