ERROR CODE
E3
:
STORAGE SENSOR FAILURE
This condition causes the following event:
¾
The Alarm Service Led is blinking
¾
The buzzer is active
¾
The display shows the ERROR code
“E3”
THIS ALARM DOSEN’T STOP THE WORKING CYCLE (CHILLING OR
FREEZING) BECAUSE THE CHECK CONTROL TEMPERATURE WILL BE
DONE BY EVAPORATOR PROBE BUT IT IS NECESSARY TO CALL THE
SERVICE to changed the probe.
ALARMS CONDITIONS
This type of error causes the following possible events:
¾
(if present)
The sterilize process is stopped and the germicidal light
turned off (if present), according to the parameter
OdS
(see
reference parameter list)
¾
The door heating element is always active
¾
During maintenance the temperature control turn to the evaporator
probe, according to positive or negative evaporator set point (PScF
or nScF).
SWICTH OFF ALARMS :
ONLY REPLACEMENT THE PROBE
- 68 -