4 | SelfCalibration
36
You can download the ColorNavigator 7 software and User's Manual from our web site:
(
Note
• When using the software, you will need to connect the external device to the monitor with the
supplied USB cable.
• For details on the USB cable connection, see
Using the Docking Station Function [
• While using the software, do not operate the power button or control buttons on the front side of
the monitor.
4.1 SelfCalibration Functions
Set the details for SelfCalibration.
Execute
SelfCalibration can be executed manually regardless of the schedule.
Note
• After selecting "Execute", warming up (after the power to the monitor is switched on, leaving the
monitor on for a certain period of time until the display becomes stable) may be executed before
the built-in calibration sensor comes out.
Settings - Mode Settings - Standard Mode
Settings: "On" / "Off"
Switch between enabling/disabling SelfCalibration in Standard Mode.
Settings - Mode Settings - Advanced Mode (CAL)
Settings: "On" / "Off"
Switch between enabling/disabling SelfCalibration in Advanced Mode.
Settings - Schedule - Start time
Settings: "Power Save" / "Immediately" / "Application" / "Off"
Select the timing to execute SelfCalibration upon reaching the time set in the schedule.
• "Power Save"
To be executed under any of the following conditions.
– When the monitor is in "Power Save" mode or power is off at the set time.
– The monitor shifts to the power saving mode or the power is turned off when the set
time in the schedule is exceeded.