4.2 LVDT Signal Conditioning Module Setup
The E9310 (2) channel and E9320 (4) channel LVDT signal conditioning modules are
signal conditioning amplifiers. The module plugs into the lower bay of the Accusetter II
and converts the outputs of the transducers into a useable format for the main controller
board.
The modules have jumpers allowing magnification reduction to be set to 10x for long
range transducers or 1x for standard transducers, see figure 4.4. Jumpers are also
available to buss signals in or out to another Accusetter II or an external device.
When an E9310 or E9320 module is used on an Accusetter II the column will
automatically recognize the setting of the 1X/10X jumper and program the column
accordingly.
Power to the column must be turned off prior to installing or removing a signal
conditioning module.
Figure 4.4 - LVDT Module E9310 or E9320
Содержание Accu-Setter II
Страница 2: ...A Accu Setter II User s Manual Edmunds GAGES...
Страница 16: ...2 3 2 3 Overall E9300Unit Figure 2 1 Accusetter II Basic Unit E9300...
Страница 18: ...2 5 Figure 2 2 Front Panel...
Страница 20: ...2 7 Figure 2 3 Rear Panel...
Страница 31: ...3 3 Front Mounting Foot Rear Mounting Bracket Figure 3 1 Base Feet Figure 3 2 Filter Regulator Mounting...
Страница 32: ...3 4 Figure 3 3 Air Connections For Air Gaging Application Only...
Страница 39: ...3 11 An example of various multipliers as related to diameter measurement follows...
Страница 40: ...3 12 The last consideration to complete the formula is the gaging function See examples below...
Страница 43: ...PROGRAMMING SINGLE CHECK MEASUREMENT 3 15...
Страница 44: ...PROGRAMMING SINGLE CHECK MEASUREMENT 3 16...
Страница 50: ...PROGRAMMING MULTIPLE CHECK MEASUREMENT 3 22...
Страница 51: ...PROGRAMMING MULTIPLE CHECK MEASUREMENT 3 23...
Страница 59: ...PROGRAMMING MULTIPLE FIXTURE MEASUREMENT 3 31...
Страница 60: ...PROGRAMMING MULTIPLE FIXTURE MEASUREMENT 3 32...
Страница 90: ...4 5 Figure 4 3 A E Block...
Страница 93: ...4 8 Figure 4 5 Input Output Jumpers...