3-47
3.5 Calibration
NOTE:
Prior to performing the automatic calibration procedure, the Air to Electric or
LVDT inputs need to be properly setup, see
Input Setup
.
Single Check Application
Two Master System
1)
Press the CAL button.
2)
Rotate the ENTER button until "MASTER" is displayed and then press ENTER. The
check name will be displayed.
3)
Press the ENTER button. "INS MAX" will be displayed.
4)
Place the MAX master into gage position and press ENTER. "GAGE MAX" will be
displayed.
5)
Press the ENTER button. The MAX master will be gaged and then the display will
update to "INS MIN".
6)
Remove the MAX master, place the MIN master into gage position and press ENTER.
"GAGE MIN" will be displayed.
7)
Press the ENTER button. The MIN master will be gaged and then the display will
update to "MASTERED" if the calibration was successful. If the calibration was
successful the gage is ready to measure parts, see
Operation
. Remove the MIN master
and press RUN . If the calibration was not completed successfully then see the
Troubleshooting section of this manual.
Zero Master Systems
1)
Press the CAL button.
2)
Rotate the ENTER button until "MASTER" is displayed and then press ENTER. The
check name will be displayed.
3)
Press the ENTER button. "INS ZERO" will be displayed.
4)
Place the ZERO master into gage position and press ENTER. "GAGE ZERO" will be
displayed.
5)
Press the ENTER button. The ZERO master will be gaged and then the display will
update to "MASTERED" if the calibration was successful. If the calibration was
successful the gage is ready to measure parts, see
Operation
. Remove the ZERO
master and press RUN . If the calibration was not completed successfully then see the
Troubleshooting section of this manual.
Содержание Accu-Setter II
Страница 2: ...A Accu Setter II User s Manual Edmunds GAGES...
Страница 16: ...2 3 2 3 Overall E9300Unit Figure 2 1 Accusetter II Basic Unit E9300...
Страница 18: ...2 5 Figure 2 2 Front Panel...
Страница 20: ...2 7 Figure 2 3 Rear Panel...
Страница 31: ...3 3 Front Mounting Foot Rear Mounting Bracket Figure 3 1 Base Feet Figure 3 2 Filter Regulator Mounting...
Страница 32: ...3 4 Figure 3 3 Air Connections For Air Gaging Application Only...
Страница 39: ...3 11 An example of various multipliers as related to diameter measurement follows...
Страница 40: ...3 12 The last consideration to complete the formula is the gaging function See examples below...
Страница 43: ...PROGRAMMING SINGLE CHECK MEASUREMENT 3 15...
Страница 44: ...PROGRAMMING SINGLE CHECK MEASUREMENT 3 16...
Страница 50: ...PROGRAMMING MULTIPLE CHECK MEASUREMENT 3 22...
Страница 51: ...PROGRAMMING MULTIPLE CHECK MEASUREMENT 3 23...
Страница 59: ...PROGRAMMING MULTIPLE FIXTURE MEASUREMENT 3 31...
Страница 60: ...PROGRAMMING MULTIPLE FIXTURE MEASUREMENT 3 32...
Страница 90: ...4 5 Figure 4 3 A E Block...
Страница 93: ...4 8 Figure 4 5 Input Output Jumpers...