4-29
A/E input not
responding
Signal conditioning card not properly installed/
Check that signal conditioning card is properly seated in lower
bay.
Incorrect air supply pressure/
Ensure the air supply to the filter/regulator assembly is at least 60
psi.
Damaged air tooling/
Check the air tooling for damaged or plug nozzles. If repair is
required contact Edmunds Gages.
Improper connection to signal conditioning module/
Ensure the air line from the tooling is securely connected to the
correct signal conditioning port and that there are no air leaks in
the line.
Mag and Zero not properly setup for the application./
Perform air check mag and zero setup procedure.
Input setup incorrectly programmed/
Check that "POL/MAG" under Xducer menu is not programmed
to 0.000 for the input & "Formula" under define is not 0.000.
Gage reads opposite of
expect value
Input polarity set incorrectly/
Check polarity settings under Xducer menu.
Bargraph reading off
scale high or low
Range set incorrectly/
Reset range to larger full scale value.
Part is out of tolerance/
If the part being gaged is out of tolerance by a large enough
amount the bargraph will read off scale high or low
Bargraph indicates out
of tolerance condition
for good part
Over/Under LIMITS not set properly/
Reprogram over and under limits.
Incorrect gage setup or damaged gage tooling/
Ensure the column is programmed correctly and that there is no
physical damage to the gage.
Bargraph not
responding in real time
Incorrect mode selected/
If the unit is programmed for TIR, +PEAK, or -PEAK the display
may not respond in real time until the RESET button is pressed.
If the unit is programmed for LIVE RDG it will always display
Содержание Accu-Setter II
Страница 2: ...A Accu Setter II User s Manual Edmunds GAGES...
Страница 16: ...2 3 2 3 Overall E9300Unit Figure 2 1 Accusetter II Basic Unit E9300...
Страница 18: ...2 5 Figure 2 2 Front Panel...
Страница 20: ...2 7 Figure 2 3 Rear Panel...
Страница 31: ...3 3 Front Mounting Foot Rear Mounting Bracket Figure 3 1 Base Feet Figure 3 2 Filter Regulator Mounting...
Страница 32: ...3 4 Figure 3 3 Air Connections For Air Gaging Application Only...
Страница 39: ...3 11 An example of various multipliers as related to diameter measurement follows...
Страница 40: ...3 12 The last consideration to complete the formula is the gaging function See examples below...
Страница 43: ...PROGRAMMING SINGLE CHECK MEASUREMENT 3 15...
Страница 44: ...PROGRAMMING SINGLE CHECK MEASUREMENT 3 16...
Страница 50: ...PROGRAMMING MULTIPLE CHECK MEASUREMENT 3 22...
Страница 51: ...PROGRAMMING MULTIPLE CHECK MEASUREMENT 3 23...
Страница 59: ...PROGRAMMING MULTIPLE FIXTURE MEASUREMENT 3 31...
Страница 60: ...PROGRAMMING MULTIPLE FIXTURE MEASUREMENT 3 32...
Страница 90: ...4 5 Figure 4 3 A E Block...
Страница 93: ...4 8 Figure 4 5 Input Output Jumpers...