2-11
2.7 E9310 (2) Channel LVDT Signal Conditioning Module
The E9310 module is a two-channel
signal conditioning
amplifier for inductive type
transducers such as
LVDT
s that converts the outputs of the transducers into a conditioned
signal for the main controller board. The module is mounted in the lower bay of the
Accusetter II.
Input magnification must be set using jumpers allowing magnification reduction to be set
to 10x for long range transducers or 1x for standard transducers.
A jumper matrix on the board allows it to accept signals in or send signals out to the
analog output connector. By placing the jumper pin for the desired signal line on one of
the six buss lines, the signal can now be sent or received by any other units connected to
the buss.
Power to the column must be turned off prior to installing or removing a signal
conditioning module.
Refer to the Advanced Operation,
LVDT Module Setup
section for additional
information on jumper settings.
Figure 2.7 - E9310 (2) Channel LVDT Signal Conditioning Module
Содержание Accu-Setter II
Страница 2: ...A Accu Setter II User s Manual Edmunds GAGES...
Страница 16: ...2 3 2 3 Overall E9300Unit Figure 2 1 Accusetter II Basic Unit E9300...
Страница 18: ...2 5 Figure 2 2 Front Panel...
Страница 20: ...2 7 Figure 2 3 Rear Panel...
Страница 31: ...3 3 Front Mounting Foot Rear Mounting Bracket Figure 3 1 Base Feet Figure 3 2 Filter Regulator Mounting...
Страница 32: ...3 4 Figure 3 3 Air Connections For Air Gaging Application Only...
Страница 39: ...3 11 An example of various multipliers as related to diameter measurement follows...
Страница 40: ...3 12 The last consideration to complete the formula is the gaging function See examples below...
Страница 43: ...PROGRAMMING SINGLE CHECK MEASUREMENT 3 15...
Страница 44: ...PROGRAMMING SINGLE CHECK MEASUREMENT 3 16...
Страница 50: ...PROGRAMMING MULTIPLE CHECK MEASUREMENT 3 22...
Страница 51: ...PROGRAMMING MULTIPLE CHECK MEASUREMENT 3 23...
Страница 59: ...PROGRAMMING MULTIPLE FIXTURE MEASUREMENT 3 31...
Страница 60: ...PROGRAMMING MULTIPLE FIXTURE MEASUREMENT 3 32...
Страница 90: ...4 5 Figure 4 3 A E Block...
Страница 93: ...4 8 Figure 4 5 Input Output Jumpers...