After storing the CO-values the next point
are the draught measurements (curser key
<down>
). For exact measuring it’s recom-
mended to reset the sensor to zero. Re-
move the draught hose off the instrument
and press
<CAL>
. The zero point of the
sensor is set now. After the draught has
stabilised it can be stored for later printout
by pressing the key
<store>
:
At the end of the measurement routine
(curser down) it is checked whether a gas
leakage at the unit could be determined.
Confirmation could be done by using the
key
<Enter>
(yes / no).
The printout (IR printer required) of all re-
sults of the measurement routine could be
done by pressing
<Print>
.
Atmosph. boiler
Draft
-0.04
hPa
Val.:
-0.04
hPa
Set zero point
CAL.
Atmosph. boiler
Gas leakage
No
Change:
↵
↵↵
↵