6. Measurement Routines (Option)
The ecom-B provides measurement routines that makes standard
checking on atmospheric boilers and fan burners easier.
6.1. Atmospheric Boilers
Select from the main menu the sub menu
„Atmosph. Boilers”
and
press
<Enter>.
The table of fuel types will be displayed. Following
fuel types can be selected:
Natural gas (B)
Town gas (B)
Coke gas (B)
Liquid gas (B)
Select with the arrow keys the correspond-
ing fuel type and confirm by pressing
<En-
ter>.
The measurement
“Draft / Pressure”
is displayed. To operate the measurement
(0 – 100 hPa) the draught hose must be
connected. In case of deviations of the
zero point it could be reset by pressing
<CAL>
.
Connect the draught hose to the instru-
ment. After the pressure has stabilised it
can be stored by pressing
<store>
for later
printout. By pressing the key
<store>
again
the value will be deleted. Using the cursor
<down>
it could be get to the next meas-
urement.
Operate the following pressure measure-
ment as described above. Not needed
measurements can be skipped by using
the cursor key
<down>
(these will not be
printed).
Type of fuel
Natural gas (B)
CO2max : 11.8
A1-fact. : 0.37
Select with
↑↓
Atmosph. boiler
Connection pressure
24.7
hPa
Val.:
24.7
hPa
Set zero point
CAL.
Atmosph. boiler
Flow pressure
21.1
hPa
Val.:
21.1
hPa
Set zero point
CAL.
Atmosph. boiler
Nozzle pres pa. load
4.4
hPa
Val.:
4.4
hPa
Set zero point
CAL.
Atmosph. boiler
Nozzle pres fu. load
13.1
hPa
Val.:
13.1
hPa
Set zero point
CAL.