Call up the menu
„Pressure Tests“
with the pre-programmed
measurement routines
„Pressure Test“
,
„Loading Test“
,
„Tight-
ness Test“
and
„Usage property“
. Parameters to each measure-
ment routine can be adjusted in the menu
„Setup“
.
Setup
Scroll with cursor keys
<Up/Down>
until the menu
“Setup”
. Press
<Enter>
to activate the menu. The following parameters can be ad-
justed for the corresponding measurement routine:
Pressure Test
- Stabilization time (1 - 10 min, default: 1 min)
- Measurement time (10 - 120 min, default: 5 min)
Loading Test
- Stabilization time (1 - 10 min, default: 1 min)
- Measurement time (10 - 120 min, default: 10 min)
- Test pressure (900 - 1200 hPa, default: 1000 hPa)
Tightness Test
- Stabilization time (1 - 10 min, default: 1 min)
- Measurement time (10 - 120 min, default: 10 min)
- Test pressure (90 - 160 hPa, default: 150 hPa)
Usage property
- Stabilization time (1 - 270 min, default: 1 min)
- Measurement time (10 - 240 min, default: 10 min)
- Test pressure (20 - 160 hPa, default: 50 hPa)
- Max. operation pressure (10 - 100 hPa, default: 23 hPa)
Each parameter can be adjusted the same way:
1. Select the parameter with cursor keys
<Up/Down>
.
2. Press
<Enter>
to activate.
3. select the digit with the keys
<Store>
(left),
<Print>
(right) and
adjust with the cursor keys
<up/down>
.
4. Confirm with
<Enter>.