XS series operating manual
XS user manual v0.55
- 123 -
*TST? :
Allows to check that the unit is correctly working.
The message is as follows : #H<NR1> with:
b3
b2
b1
b0
result
Hipot missing
X
X
X
1
#H01
Megohm missing
X
X
1
X
#H02
Continuity missing
X
1
X
X
#H04
Leakage missing
1
X
X
X
#H08
With X = 0 or 1
NOTE : the *TST? command must be used before selecting a function (from the start display)
Device commands
MEG(MEGohmmeter) :
Selects the megohmmeter. Must be sent from the initialization display.
HIP(HIPot) :
Selects the hipot tester. Must be sent from the initialization display.
LEAK
(
LEAKage
)
:
Selecs the leakage function. Must be sent from the initialization display.
GND(GrouND) :
Selects ground continuity function. Must be sent from the initialization display.
SEQ(SEQuence) :
Selects the sequence function. Must be sent from the initialization display.
CONF(CONFig) :
Selects the configuration function of the unit. Must be sent from the initialization display.
PAR(PARameter) <NR1> :
Selects the parameter set for the running function.
For instance : MEG : PAR 1
DCV(DCVoltage) <NR1> :
Sets a new value in volt to the current DC voltage parameter to the parameter set of the
running function.
For instance : DCV 500
ACV(ACVoltage) <NR1> :
Sets a new value in volt to the current AC voltage parameter to the parameter set of the
running function.
Hipot tester : Min. : 10 Max. : 5000
Leakage : Min. : 200 Max. : 270
For instance : hipot : ACV 5000 leakage : ACV 244
ACC (ACCurrent) <NR3> :
Sets a new value in ampere to the AC current parameter to the parameter set of the
ground continuity function.