XS series operating manual
XS user manual v0.55
- 51 -
Display shows :
Fig 7.4
Select the number to be modified by moving the cursor with the RIGHT/LEFT arrow keys
Increase or decrease the value with respectively the UP and the DOWN arrow keys (0 1 2 3 4 5 6 7 8
9. 0 1 2 3 etc)
Repeat operation for all the numbers if necessary
With the RIGHT arrow key, move the cursor up to the unity zone
Scroll the unity:
K
(1 000 ohms),
M
(1 000 000 ohms),
G
(1 000 000 000 ohms)
T
(1 000 000 000 000 ohms).
Enter the threshold with the ENTER key
WARNING : the HI LIMIT must be always higher than the LO LIMIT, otherwise an error message is
displayed :
HI LIMIT < LO LIMIT
To set the LO LIMIT value proceeds the same way as for the HI LIMIT.
WARNING: the LO LIMIT must be always lower than the HI LIMIT, otherwise an error message is
displayed:
LO LIMIT > HI LIMIT
Timer operation
The unit is equipped with a timer to define the insulation resistance measurement time (from 1 up
to 999 seconds). This feature is particularly interesting when measuring on capacitive specimens
of which the insulation resistance increases as a function of the measurement time. At the end of
the test time, the unit automatically stops the test and memorizes the last measured value on the
LCD screen.
In case where a test time is set to 0, the unit remains in the measurement mode permanently until
the
[DISCHARGE]
key is pressed.