![Eaton Sefelec CXS40 Скачать руководство пользователя страница 91](http://html1.mh-extra.com/html/eaton/sefelec-cxs40/sefelec-cxs40_operating-manual_526969091.webp)
XS series operating manual
XS user manual v0.55
- 90 -
Fig 11.2
Operation
A remote control accessory can be connected at any time except during the measurement time.
Inside a measurement function, the unit sends a beep to indicate the accessory connection and the
unit sends a double beep to indicate the disconnection.
A small icon on the LCD screen indicates the remote control accessory connection as well as the
accessory type ( 1 or 2 ).
Pressing on the remote control push button will start the measurement.
The push button has to be kept pressed all along the test time.
Releasing the push button will stop immediately the test.
NB : For safety reason, it is no more possible to start a test from the green front panel push
button when a remote control accessory is connected.
Fig 11.3