XS series operating manual
XS user manual v0.55
- 40 -
Connection for the ground continuity measurement
Fig 5.7
Connect the measurement probe (
CO183 or CO184
) on the UA, IA, UB, IB safety banana terminals
(GREEN terminal = CURRENT, GREY terminal = VOLTAGE).
Connect the specimen to be tested as follows:
Fig 5.8
WARNING : measurement of a specimen already connected to the earth.
When measuring a ground continuity on specimen already connected to the earth, it is mandatory
to keep the accuracy specifications, not to connect the IB and UB terminals ( label with a cross
over the earth symbol) to the specimen side connected to the earth.