XS series operating manual
XS user manual v0.55
- 127 -
Continuity
: VOLT 2.830E+00 OHM 3.230E-1 (main unit =VOLT)
B) Following characters in sequence mode
For instance :
L1 C0: 0.15m
0.00V,L2 R0: 1.50KV 0.02mA,L3 M0: 41.7 G
,L4 F0:q 0.01mA 223V
A2,L5 ..: ,L6 ..: ,L7 ..: ,L8 ..:
The syntax rules are the following :
the end of the message is : CR ( Carriage Return)
each test step starts with a «
L
» and ends with «
,
» ( except for the last one) .
the «
L
» letter is followed by a number between 1 and 8 , giving the test step number and
then by a «
space
».
then comes a letter according to the type of the test step :
Function
French
English
German
Insulation
M
M
M
Dielectric strength
R
H
H
Continuity
C
G
E
Void
. .
. .
. .
Pause
OK.
OK.
OK
the measurement type is followed by the memory number which contains the
measurement parameters , then by a separator character «
:
»
if the test step is void , the measurement will be replaced by a «
space
»
otherwise , the next character gives the test step result :
«
space
» indicates that the test is good
«
q
» indicates that the test is bad
for the insulation and dielectric strength tests the previous character is followed by 1
‘space’ and for the ground continuity test by 2 ‘space’.
then the numerical values are given . When the value cannot be measured ( no continuity
or overload ) the numerical indication is replaced by «
----
»
To check if a tested specimen is good or bad at the end of a sequence , it is necessary to
check the STB register ( bit 3) .To get the result of each test step , check if the result
character is «
space
» or «
q
».
RS232 / IEEE-488 commands summary
Code
Init
menu
setup m
Ω kV
M
Ω Seq.
REM
X
*CLS
X
X
X
X
X
X
*ESE
X
X
X
X
X
X
*ESE?
X
X
X
X
X
X
*ESR?
X
X
X
X
X
X
*IDN?
X
X
X
X
X
X
*LRN?
X
X
X
*RST
X
X
X
X
X
X
*SRE
X
X
X
X
X
X
*SRE?
X
X
X
X
X
X
*STB?
X
X
X
X
X
X