Page 30
Technical Documentation EASYRO
®
Water Treatment Systems© 2015 easymetal GmbH.
This information is subject to copyright laws and may not be reproduced or changed without express written permission of easymetal GmbH.
Clean water outlet
leads into a bucket
Fresh water supply connected
with cleaning solution in canister
VIEW OF TUBING CONNECTIONS FOR MAINTENANCE
SUCTION OF CLEANING SOLUTION
Please reinsert the power plug into the power socket.
The Display shows „CHANGE FILTER“ or no text depending on model.
Press ESC + left arrow (for 5 seconds)
= Suction for cleaning the membranes
**********************
CHANGE
FILTER
2016 - 09 - 16
**********************
EASYRO
MEMBRANE
CLEANING
PROCESS
Now the system aspirates the cleaning solution for 60 seconds.
Disconnect the power plug from the power socket as soon as all of
the cleaning solution has been aspirated and before the system starts
to suction air (= canister is almost empty).
If you use the EASYRO
®
240 or 480, please repeat the suction process 2 or 3 times until all of the
cleaning solution has been aspirated (each time, press ESC + to restart the suction process).
Allow the cleaning solution to work on the system for approx. 30 minutes.