54
Fig. 13 ATC770 Board Location
CPU BOARD
DIP SWITCH POSITIONS
Operating Mode
SW1 OFF
SW2 OFF
Calibration Mode
SW1 ON
SW2 OFF
Factory Mode
SW1 OFF
SW2 ON
Security Mode
SW1 ON
SW2 ON
8.2
G
ENERAL
C
ALIBRATION
P
ROCEDURE
1. Set the DIP switches to calibration mode as shown above. The upper display should show
CAL
while the lower display shows
ATC.
2. Use the
▼
or
▲
keys to show the following functions:
•
Firmware revision
•
Zero input counts (ZERO)
•
Pressure input counts (STR)
•
Reference junction counts (RJ)
•
Remote set point, linear temperature input and line resistance for RTD input (RSP.RL)
•
Thermocouple and RTD input (TC.RTD)
•
Digital inputs status (DIG.IN)
•
Maximum Power Consumption (normally Blank)
•
All LED’s lighted.