56
when equipped with this option, is compatible with Modbus and J-Bus protocols, the choice of
which is made in the Configuration/Setup menu.
8.3.1 S
ERIAL
C
OMMUNICATION
I
NTERFACE
A
DDRESS
This function is used to set the serial Communication Interface Address. To view or access this
function, press the
FUNC
key until
nonE and GROUP show on the display. Press the
▲
key until
3
shows in the upper display. Press the
FUNC
key and the lower display changes to
SC.ADR. The
upper display shows
OFF. Press the
▲
or
▼
keys until the upper display changes to the appropriate
address, from 1 to 255. Press the
FUNC
key to store the value, and to view the next parameter. If
finished, press
A/M
to return to the operating screen.
8.3.2 P
ROTOCOL
T
YPE
This function is used to select the Protocol Type. To view or access this function, press the
FUNC
key until
nonE and GROUP show on the display. Press the
▲
key until
3 shows in the upper display.
Press the
FUNC
key and the lower display changes to
SC.BUS. The upper display shows nodbS.
Press the
▲
or
▼
keys until the upper display changes to the appropriate protocol, either Modbus
(nodbS) or Jbus (JbuS). Press the
FUNC
key to store the value, and to view the next parameter. If
finished, press
A/M
to return to the operating screen.
8.3.3 C
OMMUNICATION
T
YPE
This function is used to select the number and format of the serial bits used in communication. To
view or access this function, press the
FUNC
key until
nonE and GROUP show on the display. Press
the
▲
key until
3 shows in the upper display. Press the
FUNC
key and the lower display changes to
SC.FRM. The upper display shows 8. Press the
▲
or
▼
keys until the upper display changes to the
appropriate bit format, 8 bit without parity
(8), 8 bit with even parity (8 E), or 8 bit with odd parity
(8 O). Press the
FUNC
key to store the value, and to view the next parameter. If finished, press
A/M
to return to the operating screen.
8.3.4 C
OMMUNICATION
B
AUD
R
ATE
This function is used to select the Communication Baud Rate. To view or access this function, press
the
FUNC
key until
nonE and GROUP show on the display. Press the
▲
key until
3 shows in the
upper display. Press the
FUNC
key and the lower display changes to
SC.BDR. The upper display
shows
19200. Press the
▲
or
▼
keys until the upper display changes to the appropriate Baud rate:
600, 1200, 2400, 4800, 9600, or 19200. Press the
FUNC
key to store the value, and to view the
next parameter. If finished, press
A/M
to return to the operating screen.
8.3.5 S
ETTING
THE
S
TATUS
OF
A
UTO
/M
ANUAL
S
ELECTION
(
IF
SUPPLIED
)
Available only if the unit has the Digital Input option.
The ATC770 Pressure/Process controller has four digital inputs that can switch between Manual and
Содержание ATC770
Страница 19: ...ATC770 Microprocessor Based Pressure Process Controller 19 Fig 2 ATC770 Outline Drawing...
Страница 23: ......
Страница 30: ...30 Fig 6 Remote Set Point Jumper Location Board E...
Страница 84: ...NOTES...
Страница 85: ...NOTES...
Страница 86: ...NOTES...
Страница 88: ...DYNISCO INSTRUMENTS 38 FORGE PARKWAY FRANKLIN MA 02038 ATTN MARKETING DEPT Place Stamp Here...