26
Dynatron® ibox™ Operator’s Manual
|
Rev. 2
|
5/14/2019
TechnicalInformation
Table 2
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The Dynatron ibox is intended for use in the electromagnetic environment specified below.
The customer or the user of the Dynatron ibox should assure that it is used in such an environment.
Emissions test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment - Guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
+/- 6 kV contact
+/- 8 kV air
Compliant
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic material,
the relative humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
+/- 2 kV for power supply line
+/- 1 kV input/output lines
Not Applicable
Mains power quality should be that of a typical
commercial or hospital environment.
Surge IEC 61000-4-5
+/- 1 kV differential mode
+/- 2 kV common mode
Not Applicable
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input lines
IEC 61000-4-11
<5 % Ut (>95 % dip in Ut) for 0,5 cycle
40 % Ut (60 % dip in Ut) for 5 cycles
70 % Ut (30 % dip in Ut) for 25 cycles
<5 % Ut (>95 % dip in Ut) for 5 seconds
Not Applicable
Mains power quality should be that of a
typical commercial or hospital environment. If
the user of the equipment requires continued
operation during power mains interruptions,
it is recommended that the equipment be
powered from an uninterruptible power
supply or a battery.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
Not Applicable
Power frequency magnetic-fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE: Ut is the a.c. mains voltage prior to application of the test level.