24
Dynatron® ibox™ Operator’s Manual
|
Rev. 2
|
5/14/2019
Technical Information
Technical
Information
The following provides technical information and guidelines for the
maintenance and care of the Dynatron ibox.
Outputs
Output Jacks:
1
Current (Intensity):
0.5 – 4.0 mA
Current Accuracy:
+/- 10%
Current Precision:
0.1 mA
Voltage at Electrodes:
+/- 0-46 Volts
Electrode Cable Jack:
Custom 2-pin polarized
Treatment Timing
Maximum Total Deliverable Dose:
3360 mA-min.
(using high quality Alkaline batteries)
Single Treatment Deliverable Dose:
1-160 mA-min.
Number of 40 mA-min. Treatments:
89
Minimum Treatment Time:
15 sec.
Maximum Treatment Time:
100 minutes
Timing Accuracy:
+/- 1 second
Dimensions
Length:
4.5”
Width:
2.66”
Height:
1.25”
Weight:
3.6 oz. (without battery or electrode cable)
Display:
Organic Light Emitting Diodes (OLED)
Environmental Conditions
Transport and Storage
a. Ambient temperature range: -10C
+50C
b. Relative humidity: 10% to 100% non-condensing
c. Atmospheric pressure: 200 hPa to 1060 hPa
Operation
a. Ambient temperature range: -10C
+40C
b. Relative humidity range: 30% to 75% non-condensing
c. Atmospheric pressure range: 700 hPa to 1060 hPa
If it is necessary to store the Dynatron ibox for an extended period
(one month or more), remove the batteries and store the device in a
cool, dry location.