Preparation
86
Connecting to DC on-board network or external battery
Note requirements for external battery, see "Technical Data",
page 118 onwards.
●
Connect external battery to device via DC battery cable S
(84 14 092) or connect DC on-board network to device via
DC on-board network cable S (84 14 048).
The plug on the connecting cable that plugs into the Savina
is coded. This enables Savina to detect whether an external
battery or a DC on-board network is connected.
●
Ensure polarity is correct.
black = –
red = +
If an external battery or DC on-board network is wrongly
connected, the device will ignore it!
●
Insert plug into DC socket on rear of device.
●
Only rechargeable batteries should be used.
The charging function of the DC power supply may cause
non-rechargeable batteries to explode during mains use.
●
Do not connect DC power supplies to the DC socket.
025 3
7
2
6
1