Operation
29
Setting apnoea time T
Apnoea
on the "Alarms" screen
page
●
Press »
Alarms
∆∆
« key. Display: screen page
»Alarms 1/1«
On the screen select the line »
T
Apnoea
« =
turn rotary knob,
activate = press rotary knob.
●
Set the desired value = turn rotary knob,
confirm = press rotary knob.
To end apnoea ventilation:
●
Press the »
Alarm Reset
« key.
Savina carries on ventilating in the original ventilation mode and
associated ventilation parameters.
During Savina settings, the status of switched on apnoea
ventilation is displayed for 6 seconds by a special information
page where, in principle, apnoea ventilation is possible.
— when changing over the ventilation mode,
— when switching on Savina,
— when reducing the frequency, if the cycle time is then
longer than T
Apnoea
.
●
To set alarm limits, see page 30