Care
68
Care
●
Observe the hospital hygiene regulations.
●
Clean and prepare the machine after each patient.
Recommendation:
●
Change the hose system and expiration valve every week.
Keep the replacement systems ready.
Dismantling
Removing parts
●
Switch off the ventilator and breathing gas humidifier,
and remove their power plugs.
1
Drain condensate from water traps and ventilation hoses.
Drain the water trap on the expiration valve.
2
Drain the water container of the humidifier.
Ventilation hoses
●
Remove ventilation hoses from the device ports.
●
Remove the water traps from the ventilation hoses.
Remove the water containers from the water traps and
the expiration valve.
●
Prepare the ventilation hoses, water traps and associated
water jars, and the Y-piece for disinfection and cleaning
by autoclaving.
3
Remove the bacterial filter and treat it in accordance with
the accompanying instructions for use.
1
09 29 1
70
1
1
2
1
1
1
0 29 1
70
3
Содержание Savina
Страница 1: ...D Dräger Medical Instructions for Use Software 2 n mt78799 Savina Intensive Care Ventilator ...
Страница 4: ...4 ...
Страница 7: ...7 Intended Use Intended Use 8 ...
Страница 16: ...16 ...
Страница 46: ...46 ...
Страница 54: ...54 ...
Страница 55: ...55 Fault Cause Remedy Fault Cause Remedy 56 ...
Страница 66: ...66 ...
Страница 109: ...109 What s what Control unit 110 Front connection block 111 Back panel 112 Abbreviations 113 Symbols 115 ...
Страница 116: ...116 ...
Страница 117: ...117 Technical data Technical data 118 ...
Страница 125: ...125 Order List Order List 126 ...
Страница 130: ...130 ...
Страница 131: ...131 ...