Fault – Cause – Remedy
57
! Apnoea Alarm off
Only in »Mask/NIV« mode (optional)
Apnoea monitoring has been switched
off.
Reset the apnoea alarm time T
Apnoe
to the desired monitoring value if
necessary.
!! Apnoea ventilation
Due to detected apnoea, Savina auto-
matically switched over to mandatory
ventilation.
Check the ventilation process.
To return to the original ventilation mode:
Press the
»Alarm Reset«
key.
!!! ASB > 4 s
Not displayed in »Mask/NIV« mode.
The ASB cycle has been switched off
3 times due to exceeding time limit.
Test ventilation system for leaks.
! ASB > 4 s
Not displayed in »Mask/NIV« mode.
The ASB cycle has been switched off due
to exceeding time limit.
Test ventilation system for leaks.
! Atmospheric pressure high
The device is being used at too high
atmospheric pressure.
Use Savina inside of the given
atmospheric pressure range,
see Technical data, page 118.
One of the pressure sensors is defective.
Due to a fault in the ambient pressure
measurement, Savina calculates the tidal
volume V
T
and minute volume MV values
on the basis of 1013 mbar, in the
absence of current measurements.
If the »Airway pressure high« alarm is
correctly set and the incorrect tidal
volume V
T
and minute volume MV values
are tolerable, Savina can continue to be
used.
Keep a check on the patient’s condition.
Call DrägerService.
! Atmospheric pressure low
The device is being used at too low
atmospheric pressure.
Use Savina inside of the given
atmospheric pressure range,
see Technical data, page 118.
One of the pressure sensors is defective.
Due to a fault in the ambient pressure
measurement, Savina calculates the tidal
volume V
T
and minute volume MV values
on the basis of 1013 mbar, in the absence
of current measurements.
If the »Airway pressure high« alarm is
correctly set and the incorrect tidal
volume V
T
and minute volume MV values
are tolerable, Savina can continue to be
used.
Keep a check on the patient’s condition.
Call DrägerService.
!!! Breathing gas temp. high
Breathing gas temperature is over 40
o
C. Switch off humidifier. Use longer
inspiration hoses.
!! Check settings
Due to an internal data loss, Savina is
operating with factory settings.
Check settings and adjust.
Message
Cause
Remedy
Содержание Savina
Страница 1: ...D Dräger Medical Instructions for Use Software 2 n mt78799 Savina Intensive Care Ventilator ...
Страница 4: ...4 ...
Страница 7: ...7 Intended Use Intended Use 8 ...
Страница 16: ...16 ...
Страница 46: ...46 ...
Страница 54: ...54 ...
Страница 55: ...55 Fault Cause Remedy Fault Cause Remedy 56 ...
Страница 66: ...66 ...
Страница 109: ...109 What s what Control unit 110 Front connection block 111 Back panel 112 Abbreviations 113 Symbols 115 ...
Страница 116: ...116 ...
Страница 117: ...117 Technical data Technical data 118 ...
Страница 125: ...125 Order List Order List 126 ...
Страница 130: ...130 ...
Страница 131: ...131 ...