Operation
22
Flow Acceleration FlowAcc
By setting the FlowAcc, the pressure and flow increase at the
start of inspiration can be changed.
To set FlowAcc:
●
Press »
Settings
∆∆
«key,
screen page »
Settings 1/1
« appears.
●
Select line »
FlowAcc
« on screen = turn rotary knob,
to confirm = press rotary knob.
●
To set value = turn rotary knob
Setting range: 5 to 200 mbar/sec
FlowAcc high: steep pressure and flow increase
●
To confirm setting = press rotary knob.
A recommended start-up setting for FlowAcc is 35 mbar/s.
IPPV can be extended using the following ventilation
parameters:
— Trigger
— Sigh
— AutoFlow (optional)
Trigger
(IPPVAssist)
To synchronize the mandatory ventilation strokes with the
patient’s efforts to breathe in. The actual frequency may be
higher than the ventilation frequency f set.
1
The green LED lights up if the Savina detects attempted
spontaneous breathing by the patient.
The trigger can be switched off if spontaneous breathing
by the patient is not expected or desired.
To set the trigger, see page 23.
1
03 3
7
388
1
08 3
7
2
6
1
Savina
1
Содержание Savina
Страница 1: ...D Dräger Medical Instructions for Use Software 2 n mt78799 Savina Intensive Care Ventilator ...
Страница 4: ...4 ...
Страница 7: ...7 Intended Use Intended Use 8 ...
Страница 16: ...16 ...
Страница 46: ...46 ...
Страница 54: ...54 ...
Страница 55: ...55 Fault Cause Remedy Fault Cause Remedy 56 ...
Страница 66: ...66 ...
Страница 109: ...109 What s what Control unit 110 Front connection block 111 Back panel 112 Abbreviations 113 Symbols 115 ...
Страница 116: ...116 ...
Страница 117: ...117 Technical data Technical data 118 ...
Страница 125: ...125 Order List Order List 126 ...
Страница 130: ...130 ...
Страница 131: ...131 ...