Care
77
Replacing the O
2
sensors
— If the following message appears:
!!! O
2
measurement inop.
Exchange O
2
sensor 1
or
!!! O
2
measurement inop.
Exchange O
2
sensor 2
or
— if calibration is no longer possible.
Savina uses one O
2
sensor for O
2
control and display
(sensor 1) and a separate O
2
sensor for O
2
monitoring
(sensor 2).
1
Swivel inspiration port downwards
2
Unscrew screw, using a coin or similar object, and remove
cover plate.
3/4
Remove used O
2
sensor from housing
3/4
Slide new O
2
sensor into respective housing for
"Sensor 1" or "Sensor 2". Turn using light pressure
until the O
2
sensor slides further into the housing.
2
Screw cover plate into place.
Sensor 1
is automatically calibrated after fitting.
Sensor 2:
●
Allow not more than 20 minutes warm-up time to elapse,
then
●
calibrate manually, see page 49.
The internal battery supplies the two O
2
sensors with current
even when Savina is switched off. This enables Savina to
supply valid O
2
measured values as soon as it is switched on.
If the internal battery is discharged, Savina does not supply any
O
2
measured values for the first 20 minutes after switching on.
The O
2
dosage operates at reduced accuracy during this time.
Using an external battery
Either 12 V batteries or 24 V batteries can be connected as
the external battery.
The use of 24 V batteries (two 12 V batteries arranged in
series) with a minimum capacity of 15 Ah each is
recommended.
With batteries of this type, the efficiency of the DC power
supply and the associated operating time is considerably
greater than with 12 V batteries of comparable capacity.
054 29 1
70
2
3
4
1
Содержание Savina
Страница 1: ...D Dräger Medical Instructions for Use Software 2 n mt78799 Savina Intensive Care Ventilator ...
Страница 4: ...4 ...
Страница 7: ...7 Intended Use Intended Use 8 ...
Страница 16: ...16 ...
Страница 46: ...46 ...
Страница 54: ...54 ...
Страница 55: ...55 Fault Cause Remedy Fault Cause Remedy 56 ...
Страница 66: ...66 ...
Страница 109: ...109 What s what Control unit 110 Front connection block 111 Back panel 112 Abbreviations 113 Symbols 115 ...
Страница 116: ...116 ...
Страница 117: ...117 Technical data Technical data 118 ...
Страница 125: ...125 Order List Order List 126 ...
Страница 130: ...130 ...
Страница 131: ...131 ...