9023927 - 2nd edition - May 2007
Page 61 of 72 Pages
The menu » Settings «
– Stack/roll
— This function can be used to set the operating mode of the Datalogger and
the Eventlogger.
●
Select the menu items » Settings «, » Datalogger «, » Set Datalogger « and
» Stack/roll « in this order, pressing the »
« key after each selection.
●
Select » Stack « or » Roll « and press the »
« key to activate it (»
«).
Clear Datalogger
— This function is used to delete all data from the Datalogger.
●
Select the menu items » Settings «, » Datalogger « and » Clear Datalogger «
in this order, pressing the »
« key after each selection.
— The message: » Clear Datalogger data « is displayed.
●
Use the »
« key to select the line » Confirm « and press the »
« key.
— The data are deleted from the Datalogger.
Clear Eventlogger
— This function is used to delete all data from the Eventlogger.
●
Select the menu items » Settings «, » Datalogger « and » Clear Eventlogger «
in this order, pressing the »
« key after each selection.
— The message: » Clear Eventlogger data « is displayed.
●
Use the »
« key to select the line » Confirm « and press the »
« key.
— The data are deleted from the Eventlogger.
Roll
when the Datalogger storage space is use up, the oldest data
will be overwritten first.
Stack
when the Datalogger storage is full, no further data can be
saved. An appropriate warning is generated.
OK
OK
OK
M
OK
OK
M
OK