9023927 - 2nd edition - May 2007
Page 53 of 72 Pages
The menu » Settings «
– Warning interval
This function is used to set the interval between the warning signals »
« on the
analogue interface.
●
Select the menu items » Settings «, » Communication «, » Analogue interface «
and » Warning interval « in this order, pressing the »
« key after each selection.
●
Use the »
« and »
« keys to set each position of the time interval (in sec-
onds) for T
1
and T
2
and press the »
« key to confirm the setting.
●
Press the »
« key to confirm the complete time interval.
– Warning level
— This function is used to set the current on the analogue interface for the warning
signal »
«.
●
Select the menu items » Settings «, » Communication «, » Analogue interface «
and » Warning level « in this order, pressing the »
« key after each selection.
●
Use the »
« and »
« keys to set each position of the current value and press
the »
« key to confirm the setting.
●
Press the »
« key to confirm the complete current value.
– Maintenance level
— This function is used to set the current on the analogue interface for the
maintenance signal »
«.
●
Select the menu items » Settings «, » Communication «, » Analogue interface «
and » Warning level « in this order, pressing the »
« key after each selection.
●
Use the »
« and »
« keys to set each position of the current value and press
the »
« key to confirm the setting.
●
Press the »
« key to confirm the complete current value.