Technical data
42
Supplement
Fabius plus SW 3.n
Device data
The following data have been added:
External fresh-gas outlet
The data for the fresh gas flow have been deleted:
General safety standards for anesthesia workstations
The following sentence has been deleted:
The following also apply for devices manufactured
from July 2014 on:
Ventilator
The accuracy of the tidal volume has been changed:
Medical gas supply through central gas supply
Average inlet flow at 2.8 kPa x 100 (40.6 psi) supply
pressure
≤
60 L/min
Power supply
Inrush current
approx. 25 to 50 A peak
The part number of the instructions for use for the power socket strip has been deleted:
The power rating cannot be configured, with optional
Dräger power socket strip (see instructions for use
for power socket strip for medical devices)
Fresh-gas flow
0 and 0.2 to 18 L/min
Accuracy
Tidal volume (
VT
)
±5 % of the set value or 20 mL, whichever value
is higher