Supplement
Fabius plus SW 3.n
19
Operation
Operation
Ventilation
The following chapters have been added or have changed:
Interdependency between the control
elements
Some settable parameters may be limited or
mutually restrict one another, so that certain
combinations of therapy settings are not possible,
e.g.,
VT
and
Freq
in the
Volume Control
ventilation mode.
If a condition is reached in which a parameter can
no longer be changed, the restricting parameter
must first be adjusted. This will enable the
adjustment of the restricted parameter.
External fresh-gas outlet with switch
and ventilator lock (option)
When the switch of the external fresh-gas outlet is
set to operation with a non-rebreathing system
, the ventilator lock prevents a controlled
ventilation mode (
Volume Control
,
Pressure
Control
,
Pressure Support
,
SIMV/PS
) from being
started.
The following alarms are issued:
From software version 3.37:
–
VENTILATOR FAIL !!!
–
CHECK APL / EXT. FGO !!!
With older software versions:
–
VENTILATOR FAIL !!!
–
CHECK APL VALVE !!!
1
Set the switch of the external fresh-gas outlet
to
COSY
.
The function of the ventilator lock is only available
with the “External fresh-gas outlet with switch and
ventilator lock” option.
This function is not available on older versions of
the external fresh-gas outlet with switch.
Ventilation mode
ManSpont
The following information has been added:
For ventilation in the
ManSpont
ventilation mode,
use a tight-fitting mask or a tube.
CAUTION
Risk of undetected fresh-gas deficiency
When the switch of the external fresh-gas outlet
is set to operation with a non-rebreathing system
[
], no alarm occurs at the start of the
ManSpont
ventilation mode.
Check that the switch of the external fresh-gas
outlet has been set to
COSY
.
41673
COSY