Supplement
Fabius plus SW 3.n
17
Assembly and preparation
Inserting a new O
2
sensor capsule
The terminology for the O2 sensor has changed.
1
Remove the O
2
sensor (A) from the inspiratory
valve.
2
Unscrew the screw cap (C) from the O
2
sensor
housing.
3
Take the new O
2
sensor capsule from the
package.
4
Place the O
2
sensor capsule (B) in the O
2
sensor housing so that the ring-shaped
conductors touch the contacts in the O
2
sensor
housing.
5
Screw on the screw cap (C) tightly by hand.
6
Insert the O
2
sensor (A) back in the inspiratory
valve.
WARNING
Risk of electric shock
If the O
2
sensor capsule is replaced during
operation, leakage currents may be
transferred.
Do not touch the patient.
20971
C
B
A
WARNING
Risk of incorrect O
2
measurement
An incorrectly fitted O
2
sensor capsule will
result in incorrect measurements.
– Make sure that the O
2
sensor capsule is
correctly inserted in the O
2
sensor
housing.
– Make sure that the O
2
sensor is correctly
inserted in the inspiratory valve.