Reprocessing
34
Supplement
Fabius plus SW 3.n
Classification of device-specific components
Observe the instructions for use for the
components.
The following classification is a recommendation
from Dräger.
Non-critical
Control elements and device surfaces including:
– Screen
– Softkeys
– Rotary knob
– O
2
flush key
– Flow control valves
– APL valve
– O
2
sensor housing
– Writing tray
– Grip bar on trolley
– Drawer handles
– Standard rails on both sides
– Clic adapter, Clic absorber
– Circuit plug on the bag elbow
– External FGO changeover switch and external
FGO port
Other surfaces which are frequently touched:
– Side parts of the housings of the screen and of
other patient monitors
– Accessory parts:
– Storage trays
– Shelf
– Hinged arms
– Probes of compressed gas hoses
– Mains plug
– Gas cylinder valves
– Transfer hose of the anesthetic gas
receiving system
– Cables and hoses that lie on floor
– Brake
Surfaces which are touched less frequently:
– Network cables and data cables
– Compressed gas hoses
– Pressure reducers
– Gas cylinders
– Drawer surfaces, outside and inside
– Anesthetic gas receiving system
– Lamp
– Holder for sample line
Semi-critical
– Breathing system:
– Breathing system housing
– Inspiratory/expiratory ports, APL valve
– Inspiratory valve, expiratory valve
– Bag elbow
– Rigid arm for breathing bag (optional)
– Flexible arm for breathing bag (optional)
– Breathing hoses
– Absorber container and absorber insert
– Soda lime dust filter (optional)
– Sample line
– When the sample line is fitted to the filter on
the Y-piece.
– When the sample line is fitted directly to the
Y-piece and the filters are fitted on the
breathing system
– Ventilator lid
– Ventilator membrane
– Ventilator hose
– Flow sensor
Critical
The device does not contain any components
classified as critical.