Supplement
Fabius plus SW 3.n
11
Overview
Wall-mounted version (optional)
The "Caution - Risk of tipping over" safety information has been
omitted and the following warning has been added in its place:
WARNING
Risk of tipping over
The wall-mounted version of the device does
not meet the requirements of ISO 80601-2-13
and IEC 60601-1 with regard to tipping
stability and transport stability. Precautionary
measures must be taken. The medical device
may tip over if handled incorrectly.
– The medical device may only be moved by
people who have the physical ability to do
so.
– Use 2 persons to transport the device.
– Make sure that there is sufficient clear
space.
– Remove any devices mounted on attached
support arms or on the top of the device.
– Fold in the support arms.
– Clear the writing tray and push it in
completely.