Reprocessing
38
Supplement
Fabius plus SW 3.n
Disinfectants
Use disinfectants that are nationally approved and
are suitable for the particular reprocessing
procedure.
Surface disinfection with cleaning
1
Remove soiling immediately. Use a
cloth dampened with disinfectant to
remove soiling.
2
Perform surface disinfection.
3
After the product has been exposed
to the disinfectant for the specified
contact time, remove residual
disinfectant.
4
Wipe with a cloth dampened with
water (at least drinking-water
quality). Allow the product to dry.
5
Check the product for visible soiling.
Repeat steps 1 to 5 if necessary.
6
Check the product for visible
damage and replace if necessary.
Manual cleaning followed by
disinfection by immersion
Manual cleaning
The cleaning agent that is used must have a pH of
between 9 and 12.
1
Prepare the cleaning solution in accordance
with the manufacturer's instructions.
2
Swirl the product backwards and forwards
several times in the solution. Make sure that
the solution reaches all surfaces and interior
spaces.
3
Rinse the product with water (at least drinking-
water quality) until residual cleaning agent is no
longer discernible.
4
Check the product for visible soiling. Repeat
steps 1 to 4 if necessary.
5
Check the product for visible damage and
replace if necessary.
Disinfection by immersion
1
Prepare the disinfectant solution in accordance
with the manufacturer's instructions.
2
Swirl the product backwards and forwards
several times in the solution. Make sure that
the solution reaches all surfaces and interior
spaces.
3
After the contact time has elapsed, rinse the
product with water (at least drinking-water
quality) until residual disinfectant is no longer
discernible.
4
Check the product for visible damage and
replace if necessary.
5
Thoroughly shake out residual water. Allow the
product to dry completely.
Machine cleaning with thermal
disinfection
Use a washer-disinfector that meets the
requirements of the standard ISO 15883. Dräger
recommends the use of a cart for anesthesia
accessories and ventilation accessories.
1
Securely position the product in the basket.
Ensure the following:
– All surfaces and interior spaces can be
flushed completely.
– The water can drain off freely.
2
Use a suitable cleaning agent.
3
Select a suitable cycle. Use demineralized
water for the final rinsing.
4
After the cycle has ended, check the product
for visible soiling. If necessary, repeat the cycle
or perform manual cleaning and disinfection by
immersion.
5
Check the product for visible damage and
replace if necessary.
6
Allow the product to dry completely.
WARNING
Risk due to penetrating liquid
Penetrating liquid may cause the following:
– Damage to the device
– Electric shock
– Device malfunctions
Ensure that no liquid penetrates the device.