Supplement
Fabius plus SW 3.n
41
Technical data
Technical data
General information
The following text has changed:
All patient-related volume values are normalized to
STAPD conditions.
All patient-related fresh-gas flow values are
normalized to STPD conditions.
Ambient conditions
The unit "psi" for ambient pressure has been added.
During operation
Temperature
10 to 35 °C (50 to 95 °F)
Ambient pressure
700 to 1060 cmH
2
O (hPa) (10.9 to 15.9 psi)
Relative humidity
20 to 80 % (without condensation)
Height
Up to 3000 m (9843 ft)
During storage and transport
Temperature
-10 to 60 °C (14 to 140 °F)
Ambient pressure
700 to 1060 cmH
2
O (hPa) (10.9 to 15.9 psi)
Relative humidity
10 to 90 % (without condensation)
The conditions for use when using additional devices can limit the environment of use of a system as a
whole. Vaporizers and anesthetic agents can limit the use of an anesthesia workstation with regard to its
temperature range and maximum fresh-gas flow. Therefore when using additional devices, follow the
associated instructions for use.