Maintenance
40
Supplement
Fabius plus SW 3.n
Maintenance
The following safety information has been added:
The following table has been amended:
The following table shows the maintenance
intervals:
NOTE
Only perform maintenance measures if there is
no patient connected to the device.
Component
Interval
Measure
Responsible personnel
CO
2
absorber
If colored violet
Replace
User
Water trap
As needed or if
soiled
Replace
User
Flow sensor
As needed or if
calibration is no
longer possible.
Clean/replace
User
O
2
sensor
As needed, if cali-
bration is no lon-
ger possible, or if
the
O
2
SENSOR
FAIL !
message is
displayed.
Replace
User
Internal lithium battery
Every 36 months
Replace
Experts
Lead-gel battery
Every 3 years
Replace
Experts
Cylinder pressure reducer for
high-pressure cylinders
1)
1) optional
After 6 years
Basic overhauling
Experts
Pressure reducer for PIN
index
1)
After 6 years
Replace
Experts