Supplement
Fabius plus SW 3.n
23
Operation
Ending operation
The following safety information has been added:
WARNING
Risk of infection
Pathogenic germs can be transmitted
between patients by the circuit plug. If a
previously used Y-piece or filter is plugged
onto the circuit plug and if a reprocessed
component is attached later in the process
(e.g., during a leakage test), the new
component may become contaminated.
Only plug reprocessed components onto the
circuit plug.