Troubleshooting
Troubleshooting
Power supply failure
Mains power supply failure
The following text has changed:
If mains power fails, Fabius automatically switches
to the internal battery.
The remaining battery charge is displayed in the
status bar.
The operating time of the battery depends on the
ventilation settings and the condition of the battery
(age and battery charge). A completely charged
battery can ensure supply for at least 45 minutes.
The following safety information has been amended:
Ventilator failure
Alarm
VENTILATOR FAIL !!!
The first step has been added:
1
Set the switch of the external fresh-gas outlet
with ventilator lock (option) to
COSY
.
2
Change to
ManSpont
ventilation mode.
3
Set the APL valve to the
Man
position.
4
Set the APL valve to the desired pressure.
5
Fill the breathing bag, if necessary with the aid
of the O
2
flush key.
6
Manually ventilate the patient.
WARNING
Insufficient ventilation of the patient
When the
BATTERY LOW !!
alarm message
(remaining battery charge 10 %) is displayed,
the ventilator will continue to operate for only
a few minutes more.
Restore mains power supply. Afterwards, the
controlled ventilation is available again.