Maintenance procedures
Zeus
Figure 5
Frame fixing screws
1. Unscrew the fixing screws holding the frame (
Figure 5
).
2. Remove the frame.
3. Replace the filter mat.
4. Check the fan for proper functioning.
5. Refit the plastic covers using the reverse method of that used for
removal.
6. Perform a functional check according to the Zeus test instructions.