Maintenance procedures
Zeus
4. Disconnect the connecting cables (
Figure 25
/2).
Figure 25
Connecting the rechargeable batteries
5. Unscrew the leads (
Figure 25
/1) and remove the rechargeable batteries.
6. Refit the rechargeable batteries in the reverse order of their removal.
7. Perform a function test of the uninterruptible power supply as per test
instructions.
CAUTION
Overheating of the rechargeable batteries.
In order to avoid overheating the rechargeable batteries during charging,
the charging cycle is controlled by a temperature sensor. To ensure proper
operation, the temperature sensor must be placed between the two
rechargeable batteries upon reassembly.
Содержание Zeus
Страница 1: ...5133 001 9036190 Revision 3 0 Because you care Technical Documentation Zeus Anesthetic workstation ...
Страница 2: ......
Страница 8: ...All rights reserved Copyright reserved K5133001IECIVZ fm 03 05 06 VI Contents ...
Страница 9: ...1 General ...
Страница 10: ...2 ...
Страница 15: ...7 Function Description ...
Страница 16: ...8 ...
Страница 85: ...77 Maintenance Procedures ...
Страница 86: ...78 ...
Страница 115: ...107 Schematics and Diagrams ...
Страница 116: ...108 ...
Страница 127: ......
Страница 128: ...119 Annex Technical Documentation acc to EMV standard IEC EN 60601 1 2 2001 Test List Spare Parts Catalogue ...
Страница 129: ...120 ...
Страница 133: ......
Страница 163: ......
Страница 165: ......
Страница 178: ......