Function Description
Zeus
The distributor block heater (4), the mixing chamber heater (88) and the satu-
rated vapour line heater (61) prevent anaesthetic condensing in the outlet of
the supply module. The heat output is regulated with the temperature sensors
(65 and 66) and the temperature sensor of the mixing chamber heater.
5.2.1
Supply Module PCB
The Supply Module PCB is located in the DIVA supply module.
The key functions of the Supply Module PCB are:
–
To control the valves of the supply module
–
To regulate the heaters of the supply module
–
Various monitoring functions
The core of the Supply Module PCB is the microcontroller 80C167CR. The
program is located in a 1 MB Flash-ROM (
Figure 16
).
Figure 16
Block diagram of the Supply Module PCB
Содержание Zeus
Страница 1: ...5133 001 9036190 Revision 3 0 Because you care Technical Documentation Zeus Anesthetic workstation ...
Страница 2: ......
Страница 8: ...All rights reserved Copyright reserved K5133001IECIVZ fm 03 05 06 VI Contents ...
Страница 9: ...1 General ...
Страница 10: ...2 ...
Страница 15: ...7 Function Description ...
Страница 16: ...8 ...
Страница 85: ...77 Maintenance Procedures ...
Страница 86: ...78 ...
Страница 115: ...107 Schematics and Diagrams ...
Страница 116: ...108 ...
Страница 127: ......
Страница 128: ...119 Annex Technical Documentation acc to EMV standard IEC EN 60601 1 2 2001 Test List Spare Parts Catalogue ...
Страница 129: ...120 ...
Страница 133: ......
Страница 163: ......
Страница 165: ......
Страница 178: ......