55
Zeus
Function Description
A working electrode and a counter-electrode are housed in an electrolyte-
filled sensor housing. The electrolyte chamber is isolated from the gas-con-
ducting section of the sensor by a diffusion barrier. At the working electrode,
which has a potential against the counter-electrode, the oxygen diffused
through the barrier is reduced. The current flow represents a measure of the
oxygen partial pressure.
At the counter-electrode, oxidation of the water in the electrolyte occurs as a
result of electrolysis. The oxygen formed at the counter-electrode is dis-
charged through a pressure compensation opening to the surrounding air.
This principle is termed an O2 pump.
No material transformation or decomposition occurs overall inside the sensor,
resulting in longer life and improved long-term stability.
Содержание Zeus
Страница 1: ...5133 001 9036190 Revision 3 0 Because you care Technical Documentation Zeus Anesthetic workstation ...
Страница 2: ......
Страница 8: ...All rights reserved Copyright reserved K5133001IECIVZ fm 03 05 06 VI Contents ...
Страница 9: ...1 General ...
Страница 10: ...2 ...
Страница 15: ...7 Function Description ...
Страница 16: ...8 ...
Страница 85: ...77 Maintenance Procedures ...
Страница 86: ...78 ...
Страница 115: ...107 Schematics and Diagrams ...
Страница 116: ...108 ...
Страница 127: ......
Страница 128: ...119 Annex Technical Documentation acc to EMV standard IEC EN 60601 1 2 2001 Test List Spare Parts Catalogue ...
Страница 129: ...120 ...
Страница 133: ......
Страница 163: ......
Страница 165: ......
Страница 178: ......