Technical Information
Technical Documentation for Zeus
®
Anaesthesia Workstation System
according to EMC standard IEC/EN 60601-1-2: 2001
General Information
The EMC conformity of the
Zeus
®
Anaesthesia Workstation System
includes the use of external cables,
transducers and accessories that are listed in the Zeus
®
List of Assessories (German: order-no. MK02818 RI
06 or greater; English: order-no. MK01159 RI 02 or greater).
Additionally, accessories may be used which do not affect EMC compliance, if no other reasons (see refer-
ence to list of accessories in the IfU) interdict the use of them. The non-observance may result in increased
emissions or decreased immunity of the equipment.
The
Zeus
®
Anaesthesia Workstation System
should not be used adjacent to or stacked with other equip-
ment; if adjacent or stacked use is inevitable, the
Zeus
®
Anaesthesia Workstation System
should be ob-
served to verify normal operation in the configuration in which it will be used.
Other equipment which can be used adjacent to or stacked with this device are listed in Zeus
®
List of Acces-
sories or if a EC Declaration of Compatibility is available by Dräger Medical.
Electromagnetic Emissions
Electromagnetic Emissions
The Zeus
®
Anaesthesia Workstation System
is intended for use in the electromagnetic environment
specified below. The user of the
Zeus
®
Anaesthesia Workstation System
should assure that is used in
such an environment.
Emissions Compliance
according to
Electromagnetic environment
RF emissions (CISPR 11)
Group 1
The
Zeus
®
Anaesthesia Workstation System
uses RF
energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
Class A
The
Zeus
®
Anaesthesia Workstation System
is suit-
able for use in all establishments other than domestic and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
(IEC 61000-3-2)
Not applicable
N/A if RF emissions are class A
Voltage fluctuations / flicker
(IEC 61000-3-3)
Not applicable
N/A if RF emissions are class A
Information re electromagnetic emissions (IEC 60101-1-2: 2001, table 201)
October, 30 2004
Содержание Zeus
Страница 1: ...5133 001 9036190 Revision 3 0 Because you care Technical Documentation Zeus Anesthetic workstation ...
Страница 2: ......
Страница 8: ...All rights reserved Copyright reserved K5133001IECIVZ fm 03 05 06 VI Contents ...
Страница 9: ...1 General ...
Страница 10: ...2 ...
Страница 15: ...7 Function Description ...
Страница 16: ...8 ...
Страница 85: ...77 Maintenance Procedures ...
Страница 86: ...78 ...
Страница 115: ...107 Schematics and Diagrams ...
Страница 116: ...108 ...
Страница 127: ......
Страница 128: ...119 Annex Technical Documentation acc to EMV standard IEC EN 60601 1 2 2001 Test List Spare Parts Catalogue ...
Страница 129: ...120 ...
Страница 133: ......
Страница 163: ......
Страница 165: ......
Страница 178: ......