Draeger Medical, Inc.
Narkomed GS Operator’s Instruction Manual
4-7
Preuse Checkout
4
b. If the system is equipped with a PEEP valve, verify the PEEP
valve's performance. Attach a breathing bag to the patient Y-piece
with an appropriate adapter such as an Draeger Medical
combination mask elbow with a 22 mm male fitting for the
breathing bag and 15 mm male fitting for the Y-piece. With the
manual/automatic selector valve in the
AUTO
position, set the
ventilator to the preferred frequency.
Set the PEEP bypass switch to the
PEEP ON
position. Then adjust
the PEEP valve to different values and observe the breathing
system pressure gauge to verify performance. Turn the PEEP valve
control knob fully counterclockwise to its lowest setting after the
test is completed. Set the PEEP bypass switch to the
PEEP OFF
position.
Monitors
10. Check the alarm limit settings. The monitor alarm limits are
automatically set to a default configuration when the
SYSTEM POWER
switch is turned on. Check these settings and adjust them if necessary.
Alarm limits can be adjusted at the beginning of or during a procedure.
Also, make sure that any external monitors (if any) are connected
properly.
11. * Test the alarm functions for all monitors. Simulate alarm conditions
and check for appropriate alarm signals.
System Flush
12. Flush the system with 100% oxygen by pressing the oxygen flush
button.
Fresh Gas
Oxygen
Sensor
Adapter
Option
Verification
13. If the optional fresh gas oxygen sensor adapter is installed, make sure
the fresh gas hose connection is intact and not occluded. Verify that the
oxygen analyzer is properly calibrated. (See “Calibrating the Oxygen
Sensor” on page 12-4 for more information.) When removing the oxygen
sensor during calibration, insert the oxygen sensor port plug into the
fresh gas adapter port.
Final Position
When the daily checkout procedure is complete, verify that:
a. all vaporizers are off (the handwheels are set to zero)
b. the APL Valve is open (fully counterclockwise)
c.
the manual/automatic switch is set to
BAG
d. all flowmeters indicate 0 (or minimum)
e. the patient suction is level adequate
f.
the breathing system is ready to use (the bag is in place and all
hoses are connected properly)
Содержание Narkomed GS
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 10: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 12: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 20: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 62: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 70: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 76: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 96: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 100: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 104: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 186: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 196: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 202: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 203: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...