3-8
Narkomed GS Operator’s Instruction Manual
Draeger Medical, Inc.
3
Daily Checkout Procedure
To perform the test:
a. Close all flow control valves on the anesthesia machine.
b. Turn the
SYSTEM POWER
switch to the
STANDBY
position.
c.
Turn the vaporizers to 0% concentration.
d. Interconnect the inspiratory valve and the expiratory hose terminal
on the ultrasonic flow sensor with a 22 mm breathing hose.
e. Set the manual/automatic selector valve to
BAG
.
f.
Close the APL valve by turning the knob fully clockwise to its stop
position.
g. Check that the breathing pressure gauge is on 0.
h. Attach the supplied test terminal to the breathing bag mount.
i.
Connect a sphygmomanometer squeeze bulb (available from
Draeger Medical) to the hose barb on the test terminal.
j.
Pump the squeeze bulb by hand until the breathing system
pressure gauge indicates pressure of at least 50 cm H
2
O (not to
exceed 80 cm H
2
O).
k. Observe the pressure drop at the breathing system pressure gauge.
When the pressure is at 50 cm H
2
O, begin counting seconds. The
pressure must not drop more than 20 cm H
2
O in 30 seconds.
Scavenger
System
25. * Verify the performance of the scavenger system.
To test the open reservoir scavenger system:
a. Make sure a 19 mm scavenger hose is connected between the
ventilator relief valve and the scavenger interface.
b. Make sure a 19 mm scavenger hose is connected between the APL
valve on the absorber and the scavenger interface.
c.
Check for moisture accumulation in the breathing and scavenger
hoses. Remove any moisture found.
d. Verify the safe performance of the open reservoir scavenging
system. With the scavenging system properly installed and
operating, test for positive and negative pressure relief.
To test for negative pressure relief:
e. Connect a vacuum hose to the DISS threaded terminal on the left-
hand side of the scavenger (or attach a wall suction hose onto the
adapter's hose-barb fitting).
f.
Short-circuit the inspiratory valve and the expiratory hose terminal
on the ultrasonic flow sensor with a 22 mm breathing hose.
Содержание Narkomed GS
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 10: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 12: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 20: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 62: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 70: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 76: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 96: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 100: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 104: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 186: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 196: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 202: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 203: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...