Draeger Medical, Inc.
Narkomed GS Operator’s Instruction Manual
2-25
General Description
2
When the hospital's emergency generator comes on-line (or when a
disconnected power cord is reconnected), the Narkomed GS automatically
switches back to AC power and recharges its battery. The battery charging
system charges the battery any time the power cord is connected to an
active AC power source. The charger can recharge a fully discharged
battery in approximately 12 hours.
If the machine is receiving AC power, but the battery voltage level is low
due to a problem with the battery charging circuit or similar hardware
malfunction, the Advisory message
RESERVE BATT LOW
appears on the
alarm display.
Machine
Functions on
Backup
Battery
Power
If the hospital's primary AC power fails, the backup battery system is
activated. All monitoring functions will continue for approximately 20
minutes if the battery is fully charged.
The following events signal activation of the backup battery system:
•
The yellow
AC POWER FAIL
indicator on the anesthesia machine's
alarm panel comes on.
•
The Caution message
AC POWER FAIL
appears on the alarm display.
•
A three-pulse pattern audio alarm sounds every 30 seconds.
When the battery reserve approaches depletion following an AC power
loss:
•
The yellow
BATTERY LOW
main switch panel indicator illuminates.
NOTE:
The
BATTERY LOW
indicator will only illuminate during an
AC power loss when battery reserves are low.
•
The Caution message
AC BATTERY FAIL
appears on the alarm display.
These alarms signify that approximately 10 minutes of backup battery
power remains.
The gas supply system remains operative. However, since the ventilator is
inoperative when battery power is cut off, you must perform manual
ventilation by bag. The machine cannot provide monitoring or alarm
functions until AC power is restored.
NOTE:
If the Narkomed GS's power cord is not plugged into an active AC
outlet for a period of 30 days or more, the backup battery may
become depleted. Plugging the power cord into an active AC outlet
for approximately 12 hours will recharge a depleted battery.
Содержание Narkomed GS
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 10: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 12: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 20: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 62: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 70: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 76: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 96: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 100: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 104: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 186: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 196: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 202: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 203: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...