Narkomed GS Operator’s Instruction Manual
Draeger Medical, Inc.
4
Preuse Checkout
To test the scavenger interface for passive systems:
a. Make sure a 19 mm scavenger hose is connected between the
ventilator relief valve and the scavenger interface.
b. Make sure a 19 mm scavenger hose is connected between the APL
valve on the absorber and the scavenger interface.
c.
Check for moisture accumulation in the breathing and scavenger
hoses. Remove any moisture found.
d. Short circuit the inspiratory valve and the expiratory hose terminal
on the ultrasonic flow sensor with a 22 mm breathing hose.
e. Set the absorber's manual/automatic selector valve to
AUTO
.
f.
If the system is equipped with a PEEP valve, turn the PEEP valve
control knob fully counterclockwise to its lowest setting.
g. Open the oxygen flow control valve to a flow of 10 l/min and occlude
the 19 mm scavenger terminal labeled
EXHAUST
.
h. After the ventilator bellows inflates, the flow of oxygen exits the
system through the positive pressure safety relief valve. At this
point, the absorber system's breathing pressure gauge must
indicate a pressure of less than 10 cm H
2
O.
Manual and
Automatic
Ventilation
Systems
9. * Test the ventilator.
a. Check for proper pressure and flow at the Y-piece during the
inspiratory and expiratory phases. Turn the
SYSTEM POWER
switch and ventilator power switch to their
ON
positions. Place the
manual/automatic selector valve in the
AUTO
position. Adjust the
oxygen flow control valve to a 3 l/min flow. Set the ventilator
frequency to 3 BPM, the I:E ratio to 1:2, and the tidal volume to
about 1 liter. (If testing the Pediatric Bellows or Adult/Pediatric
Bellows, adjust the tidal volume to 200 ml.)
Adjust the ventilator inspiratory flow control to the maximum of
the “low” zone on the flow gauge. Occlude the patient side of the
Y-piece. Fill the ventilator bellows by pressing the oxygen flush
button. Observe the breathing system pressure gauge as the
ventilator cycles.
The pressure gauge must indicate a pressure over 30 cm H
2
O when
the bellows completes its downward travel. The pressure should not
exceed 3 cm H
2
O at the end of the expiratory phase when the
bellows completes its upward travel.
Содержание Narkomed GS
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 10: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 12: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 20: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 62: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 70: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 76: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 96: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 100: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 104: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 186: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 196: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 202: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 203: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...