Breathing Pressure Monitoring
14
Setting the Threshold Pressure Alarm Limit
The threshold pressure alarm limit defines the level below which an
apneic alarm condition exists. When the patient's breathing pressure falls
below the threshold limit, a message appears in the Alarm window and an
audible alarm sounds.
The threshold pressure alarm limit is automatically set to 12 cm H
2
O at
power-up. The setting can be changed to any value from 5–30 cm H
2
O, as
long as it is less than the pressure high alarm limit. You can change the
alarm limit manually to a value you select or have the system set it
automatically to an optimum value based on the current peak pressure.
WARNING:
The threshold pressure alarm limit should be as close as
possible to the sensed peak pressure without exceeding it.
For details, see “Threshold Limit Guidelines” and “Sample
Threshold Limit Settings” in this section.
Manually
Setting the
Threshold
Limit
To set the threshold pressure alarm limit manually, follow these steps.
Remember that the value must be within the range of 5–30 cm H
2
O and
less than the pressure high alarm limit.
1. Press the Breathing Pressure Threshold key.
A box is drawn around the threshold pressure alarm limit.
2. Press the up arrow or down arrow key to increase or decrease the
threshold pressure alarm limit.
3. To save the new value, stop pressing arrow keys until the highlighting
box disappears (5 seconds), or press a different Alarm Limit key.
The new value is saved and displayed as the threshold pressure alarm
limit. If the alarm limit is on or below the upper limit of the scale, a
marker appears next to the alarm limit, and a dotted line extends from
the alarm limit horizontally across the scale.
Automatically
Setting the
Threshold
Limit
You can have the monitoring system automatically set the threshold
pressure alarm limit to an optimum value based on the current peak
pressure.
To automatically set the threshold pressure alarm limit, press and then
release the Breathing Pressure
Auto Set
key.
The LED next to the Breathing Pressure
Auto Set
key lights briefly when
you press it.
The threshold pressure limit is automatically set 4 cm H
2
O below the
current peak pressure measurement, to a minimum of 5 cm H
2
O and a
maximum of 30 cm H
2
O.
Auto
Set
Содержание Narkomed GS
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 10: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 12: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 20: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 62: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 70: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 76: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 96: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 100: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 104: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 186: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 196: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 202: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 203: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...